DimensionsSize (L×W×H) | 1250 mm × 758 mm × 680 mm |
Weight | 100 kg |
Connectivity
Data Input | USB Linked PC Integrated Barcode Reader |
Data Output | PDF, other formats available |
Compliance
Data Logged | Date/Time, Sample Description, Dispense Volume, Sample Status, Dilution Ratio, Extraction Pressure Profile, Extraction Video, User. |
Data Control | Master File Read Only Output Encrypted #-Code Validation of Data Tracking |
User Control | User Access/Login Rights |
21 CFR Part 11 | Compliant |
Method Development | Login rights Method History (User, Date/Time) Creation, Edits, Saves, Approving. |
Method Control | Level 1. – Frozen (Read Only) Approved Methods Level 2. – Edit Mode |
Sample Tracking | 2D DataMatrix Encrypted Barcode |
Process Monitoring | Pressure Monitoring, Video Monitoring |
System Monitoring | System Watchdog: Gravimetric, Pressure Monitoring, Leak Sensing |
Balance Calibration | Gravimetric (Internal Calibration) and Manual Checking |
Sample
Sample Types | Tablets (Normal and Prolonged Release), Beads, Capsules, Gels, Intermediate Granulations, Pellets, Powders, Solid & Semi-Solid Matrices, Stents, Swabs, and Liquids |
Weight | 9 mL / 14 mL |
Operational Parameters
Extraction Mode | Turbulent Flow Turbulent Flow and Ultrasonication |
Carousel Capacity | 30 Samples |
Workflow | Batch Mode up to 30 Samples or Continuous Load Mode |
Extraction Volume | 50 mL to 500 mL |
Dispensing Accuracy | ASTM Class A with Gravimetric Confirmation |
Liquid Arrival Check | Gravimetric Arrival Confirmation |
Extraction Solvent Modes | Single Solvent Extraction Mode Serial Solvent Extraction Mode |
Solvent Lines | 5 Input Lines |
Extraction & Dissolution Modes | Turbulent Flow Turbulent Flow Plus Ultrasonic Energy |
Filtrate Volume | Up to 2.4 mL |
Filter Type & Pore Size | PVDF, PTFE (Hydrophilic), Nylon, Glass Fiber, PVDF + Glass Fiber, PTFE (Hydrophilic) + Glass Fiber, Nylon + Glass Fiber. Pore Size - 0.2 μm to 1 μm |
Extracted Sample Dilution Range | 1:1 to 1:400 Standard Mode |
Flow Rate | 70 mL/min to 1 L/min |
Electrical | 100 - 240 VAC, 50/60 Hz, 1500 VA |
Sample Output
Sample Output | 20 mL Primary Vial & 2 mL HPLC Vial or Transfer Line For Online Setup |
Sample Output Volume | Up to 20 mL |
Evaporation Control | Capped and Septa Output Vials (20 mL & 2 mL) |
User Modes
Standard Mode | Quick ‘Barcode, Load & Run’ Operation Intuitive User System with Sample Progress Status and Smart Carousel Animation |
Method Development Mode | Quick ‘Barcode, Load & Run’ Operation Intuitive Method Development Setup with Full Control of Extraction and Condition Parameters |
Report & Results Mode | Report Recall and Management |
Software
Version | RES Command |
Operating System | Windows Vista Windows XP Windows 7 PRO, English Version |
Environmental
Green Design | Consumable Workflow for Reduced Solvent Consumption |