CambTEK

RES-Specitcation

CambTEK RES: Technical Specification

Dimensions
Size (L×W×H) 1250 mm × 758 mm × 680 mm
Weight 100 kg

Connectivity
Data Input USB Linked PC
Integrated Barcode Reader
Data Output PDF, other formats available

Compliance
Data Logged Date/Time, Sample Description, Dispense Volume, Sample Status, Dilution Ratio, Extraction Pressure Profile, Extraction Video, User.
Data Control Master File Read Only Output
Encrypted #-Code Validation of Data Tracking
User Control User Access/Login Rights
21 CFR Part 11 Compliant
Method Development Login rights
Method History (User, Date/Time) Creation, Edits, Saves, Approving.
Method Control Level 1. – Frozen (Read Only) Approved Methods
Level 2. – Edit Mode
Sample Tracking 2D DataMatrix Encrypted Barcode
Process Monitoring Pressure Monitoring, Video Monitoring
System Monitoring System Watchdog:
Gravimetric, Pressure Monitoring, Leak Sensing
Balance Calibration Gravimetric (Internal Calibration) and Manual Checking

Sample
Sample Types Tablets (Normal and Prolonged Release), Beads, Capsules, Gels, Intermediate Granulations, Pellets, Powders, Solid & Semi-Solid Matrices, Stents, Swabs, and Liquids
Weight 9 mL / 14 mL

Operational Parameters
Extraction Mode Turbulent Flow
Turbulent Flow and Ultrasonication
Carousel Capacity 30 Samples
Workflow Batch Mode up to 30 Samples or Continuous Load Mode
Extraction Volume 50 mL to 500 mL
Dispensing Accuracy ASTM Class A with Gravimetric Confirmation
Liquid Arrival Check Gravimetric Arrival Confirmation
Extraction Solvent Modes Single Solvent Extraction Mode
Serial Solvent Extraction Mode
Solvent Lines 5 Input Lines
Extraction & Dissolution Modes Turbulent Flow
Turbulent Flow Plus Ultrasonic Energy
Filtrate Volume Up to 2.4 mL
Filter Type & Pore Size PVDF, PTFE (Hydrophilic), Nylon, Glass Fiber, PVDF + Glass Fiber, PTFE (Hydrophilic) + Glass Fiber, Nylon + Glass Fiber.
Pore Size - 0.2 μm to 1 μm
Extracted Sample Dilution Range 1:1 to 1:400 Standard Mode
Flow Rate 70 mL/min to 1 L/min
Electrical 100 - 240 VAC, 50/60 Hz, 1500 VA

Sample Output
Sample Output 20 mL Primary Vial & 2 mL HPLC Vial or Transfer Line For Online Setup
Sample Output Volume Up to 20 mL
Evaporation Control Capped and Septa Output Vials (20 mL & 2 mL)

User Modes
Standard Mode Quick ‘Barcode, Load & Run’ Operation
Intuitive User System with Sample Progress Status and Smart Carousel Animation
Method Development Mode Quick ‘Barcode, Load & Run’ Operation
Intuitive Method Development Setup with Full Control of Extraction and Condition Parameters
Report & Results Mode Report Recall and Management

Software
Version RES Command
Operating System Windows Vista
Windows XP
Windows 7 PRO, English Version

Environmental
Green Design Consumable Workflow for Reduced Solvent Consumption